• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION TUBE SET, STD VOL, STER; SYSTEM/DEVICE, PHARMACY COMPOUNDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION TUBE SET, STD VOL, STER; SYSTEM/DEVICE, PHARMACY COMPOUNDING Back to Search Results
Catalog Number H93811
Device Problems Disconnection (1171); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/18/2021
Event Type  malfunction  
Event Description
It was reported that a sterile repeater pump tube set leaked.A disconnection was observed between the transfer tube set and the blue inlet connector (outlet connector).This issue occurred after completion of the filling process, when removing the device.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Additional information was added to d9, h3, h4, h6 and h10.H4: the lot was manufactured from august 24, 2020 - august 25, 2020.H10: the device was received for evaluation.Visual inspection was performed which observed that the tube set pvc (polyvinyl chloride) was detached from pump head assembly connector.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUBE SET, STD VOL, STER
Type of Device
SYSTEM/DEVICE, PHARMACY COMPOUNDING
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt. 001 mz. 105
c. industrial lt. 001 mz. 105
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key13025849
MDR Text Key282961796
Report Number1416980-2021-07414
Device Sequence Number1
Product Code NEP
UDI-Device Identifier00085412475721
UDI-Public(01)00085412475721
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2023
Device Catalogue NumberH93811
Device Lot Number60254791
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/18/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NA.
-
-