Model Number EL5ML
Device Problem Failure to Form Staple (2579)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/16/2021
Event Type  malfunction  
Event Description
It was reported that during a laparoscopic hysterectomy, clips were formed in the teardrop shape.The device was used on the blood vessel.No bleeding or damage of the target tissue was observed.Another device was used to complete the case.There were no adverse consequences to the patient.The patient¿s condition is stable.
 
Manufacturer Narrative
(b)(4).Batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
(b)(4).Date sent: 1/28/2022.D4: batch # v9599a.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Upon visual analysis of the returned sample, it was determined the el5ml device was returned with no apparent damage.In an attempt to replicate the reported incident, the instrument was tested for functionality.During the analysis, the device was cycled and it fed and formed one conforming clip.The event described could not be confirmed as the device performed without any difficulties noted and no product defect was identified.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following: "caution: do not excessively twist or torque the instrument jaws when positioning or firing the instrument on a tubular structure or vessel.Excessive twisting or torquing may result in clip malformation.Do not insert the clip applier through a trocar if a clip is present in the jaws.This may result in clip malformation, dislodged clips, or damage to the instrument.If a clip is present in the jaws, fully squeeze the trigger against the handle, then fully release the trigger to release the clip from the jaws before inserting the device through the trocar." as part of our quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 
6107428552
MDR Report Key13032238
MDR Text Key282616464
Report Number3005075853-2021-07929
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/17/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2022
Is the Reporter a Health Professional? No
Date Manufacturer Received01/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/03/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-