It was reported that during a hysterectomy procedure, the device got damaged.The tip of the device was broken after using and "pins" were jammed in it.No fragments were generated, surgery was successfully completed, and there was no patient consequences.
|
(b)(4).Date sent: 6/30/2022.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that the el5ml found that the device was received with one of the jaws protruding from its normal position.Upon cycling, the instrument was noted to be locked out.The device was disassembled, and one of the jaws was noted to be broken at the bifurcation area, evidence of corrosion was found throughout the broken area.Also, no clips were found on the clip track, the device was received empty.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.The event reported was confirmed and it is related to improper use of the device.The most likely reason for jaw bifurcation breakage is stress corrosion cracking and the most likely root cause is exposure to a solution containing chlorine.The reported complaint was confirmed.In addition, in order to avoid this kind of issues please do not reuse, reprocess, or re-sterilize the device.Reuse, reprocessing or re sterilization may compromise the structural integrity of the device and/or lead to device failure that in turn may result in patient injury, illness or death.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.
|