• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE Back to Search Results
Model Number GF-UCT180
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
Olympus medical systems corp.(omsc) was informed that during the incoming inspection for repair at the olympus local service department on (b)(6) 2021, it was found that the foreign material adhering to the forceps elevator and the nozzle.Other detailed information was not provided.There was no report of patient injury associated with the event.
 
Manufacturer Narrative
The subject device was returned to the olympus local service department.A supplemental report will be submitted, if additional or significant information becomes available at a later time.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Device history record review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.The exact cause of this event could not be conclusively determined.Olympus medical systems corp.(omsc) surmised that the reported phenomenon occurred due to the following cause.- the foreign material adhered to the forceps elevator and the nozzle during the procedure, but the foreign material was not removed by reprocessing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE
Type of Device
ULTRASOUND GASTROVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13041957
MDR Text Key282513089
Report Number8010047-2021-16397
Device Sequence Number1
Product Code ODG
UDI-Device Identifier04953170356346
UDI-Public04953170356346
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K093395
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberGF-UCT180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/24/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/25/2021
Initial Date FDA Received12/20/2021
Supplement Dates Manufacturer Received01/24/2022
Supplement Dates FDA Received02/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/05/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-