Model Number 1001.001
Device Problem Device Slipped (1584)
Patient Problem Laceration(s) (1946)
Event Date 12/01/2021
Event Type  Injury  
Manufacturer Narrative
Due to the nature of the deviation it is unlikely that the device contributed to the event.We suspect, that maybe the pinning technique has been not optimal as described in the instruction manual: "adjust the skull clamp to the width of the patient's head in the manner that the two skull pins in the rocker arm are equidistant from the centerline of the head and the single skull pin at the extension assembly is in line with this centerline.".
 
Event Description
Customer informed our service department on the 2nd of december that one of our products was involved in a case where a laceration occured.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORO® QR3 SKULL CLAMP
Type of Device
DORO® QR3 SKULL CLAMP
Manufacturer (Section D)
PRO MED INSTRUMENTS GMBH
boetzinger str. 86
freiburg, baden-wuerttemberg 79111
GM  79111
Manufacturer (Section G)
PRO MED INSTRUMENTS GMBH
boetzinger str. 86
freiburg, baden-wuerttemberg 79111
GM   79111
Manufacturer Contact
christopher schmitz
boetzinger str. 86
freiburg, baden-wuerttemberg 79111
GM   79111
MDR Report Key13043122
MDR Text Key282499805
Report Number3003923584-2021-00023
Device Sequence Number1
Product Code HBL
UDI-Device Identifier04250435506196
UDI-Public04250435506196
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001808
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 12/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number1001.001
Device Catalogue Number1001.001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/02/2021
Initial Date FDA Received12/20/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/14/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-