• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS; GAUGE, DEPTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS; GAUGE, DEPTH Back to Search Results
Catalog Number 319.006
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/22/2021
Event Type  malfunction  
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2021, it was noticed that old style gauges found in distal radius/mod mini frag lcp/va midfoot set, etc.Are functionally bend, break, and become loose easily and have a short lifespan.There was no patient harm or procedure reported.This complaint involves five (5) devices.This report is for (1) depth gauge for 2.0mm and 2.4mm screws.This report is 5 of 5 for (b)(4).
 
Manufacturer Narrative
Product complaint # (b)(4).Additional narrative: complainant part is not expected to be returned for manufacturer review/ investigation.Without a lot number, the device history records review could not be completed as no product was received.The investigation could not be completed, no product was received; no conclusion could be drawn at the time of filing this report.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPTH GAUGE FOR 2.0MM AND 2.4MM SCREWS
Type of Device
GAUGE, DEPTH
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key13044098
MDR Text Key287258238
Report Number2939274-2021-07162
Device Sequence Number1
Product Code HTJ
UDI-Device Identifier07611819707238
UDI-Public(01)07611819707238
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number319.006
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/22/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-