U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION JETSTREAM XC; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION JETSTREAM XC; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Model Number 45007
Device Problems Suction Problem (2170); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/06/2021
Event Type  malfunction  
Event Description
It was reported that the device had loss of aspiration.A jetstream xc 2.1mm catheter was selected for use in a superficial femoral artery (sfa) atherectomy as treatment for peripheral artery disease.This device made three successful passes though the lesion, however after the third pass there was a loss of aspiration and clicking noises were observed.The catheter was removed from the patient and the console was powered down and restarted, however the second attempt with the same device was also unsuccessful at gaining aspiration.Another jetstream device was used successfully to complete this procedure without sequela.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JETSTREAM XC
Type of Device
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road cork
cork IRELA ND
EI   IRELAND
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13044656
MDR Text Key282558934
Report Number2134265-2021-15789
Device Sequence Number1
Product Code MCW
UDI-Device Identifier08714729889892
UDI-Public08714729889892
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/23/2023
Device Model Number45007
Device Catalogue Number45007
Device Lot Number0026848465
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/06/2021
Initial Date FDA Received12/20/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/22/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-