OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE Back to Search Results
Model Number GF-UCT180
Device Problem Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2021
Event Type  malfunction  
Event Description
Customer reported image issue, blurry image, crystals were broken.The issue found during an unspecified procedure.There was no patient impact or harm reported due to the event.No user injury reported.Device return evaluation found the c-body (bending section) forceps cover glue peeling.
 
Manufacturer Narrative
The subject device was received and evaluated.Inspection found multiple broken elements on the um (ultrasound image).The customer reported issue was confirmed.Furthermore, peeling was observed on the c-body (bending section) forceps cover glue and crack was noted on the a-rubber.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 9 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, the root cause of the adhesive peeling on the distal end could not be identified olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE
Type of Device
ULTRASOUND GASTROVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13047926
MDR Text Key288589094
Report Number8010047-2021-16432
Device Sequence Number1
Product Code ODG
UDI-Device Identifier04953170356339
UDI-Public04953170356339
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K093395
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGF-UCT180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/26/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/06/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-