MAUDE Adverse Event Report: VERAN MEDICAL TECHNOLOGIES, INC SPIN DRIVE SOFTWARE APPLICATION; SFT-0011

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERAN MEDICAL TECHNOLOGIES, INC SPIN DRIVE SOFTWARE APPLICATION; SFT-0011 Back to Search Results
Model Number SFT-0011
Device Problems Computer Software Problem (1112); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/18/2021
Event Type  Injury  
Event Description
The rep and the physician had a tough time getting registration to line up.The collected 4 point clouds and the accuracy was not up to the doctor's standards.The area of interest was on the edge of the secondary carina sphere and the doctor wanted it to be more within the circle more.The rep noticed there was a lot of motion in the expiration scan.The rep uploaded the inspiration scan as a referral and collected a 5th cloud.Registration looked more accurate after this so the doctor moved forward with the procedure and into navigation the doctor tried to navigate out of the patient, but was unable to reach the nodule.The doctor completed the procedure with a bal.There was a delay in procedure of 30 minutes.
 
Event Description
Additional information regarding the reported event: this event was reported due to the 30-minute procedure delay while the patient was under anesthesia.
 
Manufacturer Narrative
This report is being supplemented to provide information that was inadvertently not provided in the initial medwatch report and to provide the results of the investigation of the event.The procedure playback and data were reviewed.Based on the review, the point cloud collected did not match the computed tomography (ct) scan that was collected, demonstrating that there was anatomical shift between the time of the ct scan and the navigated procedure.The software correctly reported a registration error and warned the user.The investigation determined that the software functioned as intended.The user attempted to register and navigate for approximately 30 minutes before determing they were unable to reach the target.No evidence of a device malfunction.Veran will continue to monitor field performance for this device.This supplemental report is being submitted as a corrective action following a retrospective review of complaints in response to an observation from an external inspection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPIN DRIVE SOFTWARE APPLICATION
Type of Device
SFT-0011
Manufacturer (Section D)
VERAN MEDICAL TECHNOLOGIES, INC
1938 innerbelt business center
saint louis MO 63114
Manufacturer (Section G)
VERAN MEDICAL TECHNOLOGIES, INC
1938 innerbelt business center
saint louis MO 63114
Manufacturer Contact
jennifer andre
1938 innerbelt business center
saint louis, MO 63114
3146598500
MDR Report Key13048702
MDR Text Key282548719
Report Number3007222345-2021-00023
Device Sequence Number1
Product Code JAK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K170023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberSFT-0011
Device Catalogue NumberSFT-0011
Device Lot Number4.3.1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/18/2021
Initial Date FDA Received12/20/2021
Supplement Dates Manufacturer Received02/08/2022
Supplement Dates FDA Received08/03/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-