As reported, image issue, black spots/dots, broken crystals were observed on the device.The issue was found during an unknown procedure.There was no patient impact or harm reported due to the event.No user injury reported.Device evaluation found forceps cover glue peeling, probe unit loose.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 9 years since the subject device was manufactured.Based on the results of the investigation, there was scratches around the area pointed out.We, therefore, cannot rule out the possibility that some kind of external force was applied to the relevant part.The following is stated in the instructions for use which can detect the event: "inspect the external surface of the entire insertion section including the bending section and the distal end for dents, bulges, swelling, scratches, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities." olympus will continue to monitor field performance for this device.
|