MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE Back to Search Results
Model Number GF-UCT180
Device Problem Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2021
Event Type  malfunction  
Manufacturer Narrative
The subject device was received and evaluated.Device evaluation found the reported customer issue was able to be duplicated as six broken items within a 14 elements gap found on um (ultrasound image) unit.Probe unit is loose likely affecting the u-image.Further inspecting found crack on a-rubber and peeling on c-body (bending section) forceps cover glue observed.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Event Description
As reported, image issue, black spots/dots, broken crystals were observed on the device.The issue was found during an unknown procedure.There was no patient impact or harm reported due to the event.No user injury reported.Device evaluation found forceps cover glue peeling, probe unit loose.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 9 years since the subject device was manufactured.Based on the results of the investigation, there was scratches around the area pointed out.We, therefore, cannot rule out the possibility that some kind of external force was applied to the relevant part.The following is stated in the instructions for use which can detect the event: "inspect the external surface of the entire insertion section including the bending section and the distal end for dents, bulges, swelling, scratches, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE
Type of Device
ULTRASOUND GASTROVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13049692
MDR Text Key284452099
Report Number8010047-2021-16469
Device Sequence Number1
Product Code ODG
UDI-Device Identifier04953170356339
UDI-Public04953170356339
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K093395
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGF-UCT180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/25/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/29/2021
Initial Date FDA Received12/20/2021
Supplement Dates Manufacturer Received01/24/2022
Supplement Dates FDA Received02/09/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/25/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-