ZIMMER SURGICAL, INC. DERMATOME AN HANDPIECE Back to Search Results
Model Number N/A
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Failure to Cut (2587)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/24/2021
Event Type  malfunction  
Event Description
It was reported that the graft was thicker than expected.The event occurred during surgery.There was no report of harm or delay.No adverse event was reported as a result of this malfunction.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is completed, a follow-up/final report will be submitted.
 
Manufacturer Narrative
This complaint has been recorded under (b)(4).Review of the most recent repair record identified no repairs related to the reported event.The rpm's and calibration were in specification and the position of the control bar was correct.A definitive root cause cannot be determined.The reported event could not be duplicated.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DERMATOME AN HANDPIECE
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key13051287
MDR Text Key282555949
Report Number0001526350-2021-01370
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number88710100
Device Lot Number63578135
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/02/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-