Model Number AAS00161-20 |
Device Problems
Device Displays Incorrect Message (2591); Application Program Problem (2880); Connection Problem (2900)
|
Patient Problem
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 12/06/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during the procedure, the system was showing a locatable sub-system (lss) disconnected error.Tried to restart the system a couple of times but with no luck.Verified the internet protocol (ip) address on the tracker and it was correct, power was on all components, removed and re plugged the orange ethernet cable, and verified that the lss files were all there and there were no issues found in the device manager.Attempted a different ethernet cable and no fix.The procedure was not completed and was not completed by other means.The patient was under anesthesia.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during the procedure, the system was showing a locatable sub-system (lss) disconnected error.Tried to restart the system a couple of times but with no luck.Verified the internet protocol (ip) address on the tracker and it was correct, power was on all components, removed and re plugged the orange ethernet cable, and verified that the lss files were all there and there were no issues found in the device manager.Attempted a different ethernet cable and no fix.The procedure was not completed and was not completed by other means.The patient was under anesthesia.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during the procedure, the system was showing a "locatable sub-system (lss) disconnected error".Tried to restart the system a couple of times but with no luck.Verified the internet protocol (ip) address on the tracker and it was correct, power was on all components, removed and re plugged the orange ethernet cable, and verified that the lss files were all there and there were no issues found in the device manager.Attempted a different ethernet cable and no fix.The procedure was not completed and was not completed by other means.The patient was under anesthesia.
|
|
Manufacturer Narrative
|
Additional information: e1(last name and phone number of initial reporter), g3, h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Additional information: g3, h3 out of an abundance of caution, superdimension is filing this mdr due to the additional risk associated with multiple exposures to general anesthesia.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the unit had a blown fuse.The issue was resolved by replacing the location subsystem, console label set, locatable guide connecting cable and location board.It was reported that the patient procedure was cancelled due to a system issue and a system error code was displayed.The system did not detect one of the system components.The reported issues were confirmed.The most likely cause was traced to a component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|