The evaluation found a portion of the adhesive at the distal end forceps cover was peeled or cracked off.The customer's reported issue was confirmed as the scope was leaking from the instrument channel.Additionally, the bending section cover glue is cracked and the elevator channel inlet is loose.A review of the scope's instrument history showed no previous repair records.The investigation is ongoing; therefore, the root cause of the reported issue cannot be determined at this time.However, if additional information becomes available a follow up medical device report will be supplemented accordingly.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been less than one (1) year since the subject device was manufactured.Based on the results of the investigation, regarding the reported issue forceps cover glue peeling, there are scratches around the indicated part, and we cannot deny that some external force was applied to the part.As a result of checking the condition of the applicable part there was no adhesive applied where it should be normally.The following is stated in the instructions for use which may detect the event: "inspect the external surface of the entire insertion section including the bending section and the distal end for dents, bulges, swelling, scratches, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities." olympus will continue to monitor field performance for this device.
|