Model Number 71440044
Device Problems Break (1069); Activation, Positioning or Separation Problem (2906)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/02/2021
Event Type  malfunction  
Manufacturer Narrative
Internal reference number: (b)(4).
 
Event Description
It was reported that, the locking bearing of the gii quick connect handle has a chip in it preventing it from locking into tibial trial.As this was notice in a field inspection, not patient was involved.
 
Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.The manufacturing process of the device did not contribute to the reported event.A visual inspection of the returned device confirmed the stated failure mode.The device has a chip on the tip of the ball bearing making it difficult to hold the material in place, rendering the device inoperative.The device shows signs of extensive use.A review of complaint history revealed similar events for the listed device, but no similar events for the batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.Additional information; d4 and d4/h5.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII QUICK CONNECT HANDLE
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin
austin, TX 78735
5123913905
MDR Report Key13055023
MDR Text Key284734703
Report Number1020279-2021-08767
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010213860
UDI-Public03596010213860
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71440044
Device Catalogue Number71440044
Device Lot Number14BM11891
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2021
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/02/2021
Initial Date FDA Received12/21/2021
Supplement Dates Manufacturer Received01/25/2022
Supplement Dates FDA Received01/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/18/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-