According to the reporter, on a laparoscopic cholecystectomy, while the device was being applied in the common bile duct (cbd), the clips were malformed.Another device was used to complete the case.There was no patient injury.
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted that malformed clips were provided.Functionally, the instrument was applied to appropriate test media, and cycled without binding.Clips advanced into the jaws, formed properly, and were held securely in place after full formation was achieved and the firing handle was released.When the cartridge was empty, the interlock engaged to prevent the jaws from approximating.It was reported that the clips did not close completely and were malformed.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur if the jaws are twisted excessively or tissue is manipulated with the jaws when firing the instrument.Deflecting the jaws and/or shaft during firing may result in an improperly formed clip and possible bleeding and/or leakage.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: when firing the instrument, squeeze the handle as far as it will go.Failure to squeeze the handle completely may result in clip malformation and possible bleeding and or leakage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|