Model Number 1458Q/86
Device Problems Product Quality Problem (1506); Failure to Sense (1559); Fitting Problem (2183); No Pacing (3268)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/08/2021
Event Type  malfunction  
Event Description
It was reported that during an implant procedure, a left ventricular lead was unable to sense or pace.This was later attributed to be due to damaged pacing system analyzer (psa) cables.It was also noted that the lead had problems with fitting the connector sleeve on its body.The lead was not used and a different lead was implanted.There were no patient consequences.
 
Manufacturer Narrative
The reported events of lead would not fit into the is-4 connector sleeve, no pacing and failure to sense were confirmed.As received, a complete lead was returned for evaluation.Electrical tests did not find any shorts or discontinues on any of the conduction paths.The lead connector passed insertion testing into the test icd device but failed to insert into the test is-4 connector sleeve.Dimensional analysis of the connector identified an over-sized diameter in a section of the connector boot.This may have contributed to the reported field events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key13056228
MDR Text Key282614909
Report Number2017865-2021-39535
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 02/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberA000111741
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/24/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATACHY RIGHT VENTRICULAR LEAD; PACING SYSTEM ANALYZER; QUADRA ASSURA; TENDRIL RIGHT ATRIAL LEAD
Patient Age72 YR
Patient SexFemale
-
-