ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG COBAS® LIAT® SARS-COV-2 & INFLUENZA A/B; COVID-19 MULTI-ANALYTE RESPIRATORY PANEL NUCLEIC ACID DEVICES Back to Search Results
Catalog Number 09211101190
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Output Problem (3005); Non Reproducible Results (4029)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/21/2021
Event Type  malfunction  
Manufacturer Narrative
Data was provided and reviewed and it was observed that run (b)(6) produced a positive sars result with a ct of 37.1.This ct value indicates very low level virus present and the curve does not appear to have any abnormalities.Overall, nothing was seen to indicate the analyzer may be performing sub-optimally.Hence the data set only included 6 runs, indication that the sample contained a viral target concentration at the limit of detection of the assay.Therefore, discrepancies at this level can be expected with repeat testing.The issue is sample specific and no product problem was identified.(b)(4).
 
Event Description
In light of the covid-19 pandemic and the subsequent emergency use authorizations (euas) for sars-cov-2 diagnostic tests, the agency has requested heightened reporting beyond the reasonably suggests requirements of 803 to include allegations of false positive or false negative results independent of harm or malfunction or off-label use.Pursuant to the agency¿s instruction, we hereby submit this mdr.A customer alleged a discrepant result generated with the cobas® sars-cov-2 & influenza a/b nucleic (scfa) acid test for use on the cobas® liat® system.The customer reported that the initial test generated sars-cov-2 positive.The same sample was repeated with the genexpert cepheid generated negative results for all 3 targets.The negative results were reported to the patient and or/ medical personnel treating the patient.No apparent harm or injury occurred in relation to the event.The customer collected a deep throat saliva in swen brand vtm sample, which is considered an off label practice.The method sheet indicates that ; this test is intended to be used for the detection of sars-cov-2, influenza a and influenza b rna in nasal and nasopharyngeal swab samples collected in a copan utm-rt system (utm-rt) or bd¿ universal viral transport system (uvt) or thermo fisher¿ scientific remel¿ media, or thomas scientific mantacc¿ premeasured 3 ml 0.9% physiological saline solution.Testing of other sample or media types may lead to inaccurate results.An investigation was conducted to evaluate the customer¿s allegation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS® LIAT® SARS-COV-2 & INFLUENZA A/B
Type of Device
COVID-19 MULTI-ANALYTE RESPIRATORY PANEL NUCLEIC ACID DEVICES
Manufacturer (Section D)
ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
1080 us highway 202 south
branchburg NJ 08876
Manufacturer (Section G)
ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
1080 us highway 202 south
na
branchburg NJ 08876
Manufacturer Contact
stacie-ann creighton
1080 us hwy 202 s
na
branchburg, NJ 08876
9082537112
MDR Report Key13056256
MDR Text Key285843546
Report Number2243471-2021-03870
Device Sequence Number1
Product Code QLT
Combination Product (y/n)N
Reporter Country CodeHK
PMA/PMN Number
EUA201779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2022
Device Catalogue Number09211101190
Device Lot Number10802Y
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/22/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Patient Sequence Number1
Patient Age45 YR
Patient SexFemale
-
-