MAUDE Adverse Event Report: GETINGE DISINFECTION AB 86-SERIES; DISINFECTOR, MEDICAL DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GETINGE DISINFECTION AB 86-SERIES; DISINFECTOR, MEDICAL DEVICES Back to Search Results
Model Number 8666
Device Problems Failure to Disinfect (1175); Leak/Splash (1354); Failure to Pump (1502); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/29/2021
Event Type  malfunction  
Event Description
On (b)(6) 2021 getinge became aware of an issue with the 86-series washer disinfector with the model name 8666.As it was stated the water and the washing detergent were leaking out of detergent pump cover.The leakage was greatest in the washing phase.Tha gathered information provides that the customer was using the washer disinfector despite the sever leakage.We also do not have any information if and how often the user checks the cycle results.During the troubleshooting the unit was repaired and returned to the service.There was no injury or damage reported however we decided to report the issue based on the potential as the issue may have negative impact at the final result of the cleaning process and the uncleaned goods may have been used to the patients¿ treatment.
 
Manufacturer Narrative
Additional information will be provided following conclusion of the investigation.Device not returned to the manufacturer.
 
Manufacturer Narrative
According to the reporting timeframe we would like to provide the information about current status of the issue.Please be advised that it is being investigated.Additional information will be provided following the conclusion of the investigation.
 
Event Description
Manufacturer's reference number: (b)(4).
 
Manufacturer Narrative
On november 29th, 2021, getinge became aware of an issue with washer disinfector model name 8666 with serial number (b)(6) and catalog number s-8666913-ctom.The device was manufactured on november 29th, 2010 and it was installed at the facility on december 08th, 2011.As it was stated in complaint record, the washing detergent and water was leaking out of the detergent pump cover.The customer is using one of washing detergents neutrawash plus or tec wash iii, which are used to provide effective cleaning during washing phase.The gathered information provides that the issue could cause incorrect dosing of washing detergent during the process and the customer was using the washer disinfector despite the severe leakage.We are not aware of any adverse event in relation to the incident, however we decided to report the issue based on the potential that any non-properly cleaned goods may have been used for the patient treatment and may have become a source of cross-infection.During initial inspection getinge technician detected that detergent pump tubing was worn and he has replaced defective squeeze tubes in all three pumps on the device.He has tested the operation of the pumps, evaluated as satisfactory and put the device back in service.The review of reportable events registered for 86-series devices manufactured by getinge disinfection ab reported to company¿s complaint handling system was performed.We were able to establish that this is one of several reported cases related to possible incorrect washing detergent dosing on this type of devices reported within the last 5 years and it is not a part of increasing trend.At the time of the event occurrence the unit was directly involved and failed to meet its specifications.However, none of the provided information indicates that upon the event occurrence the device was being used for patient treatment or diagnosis.Review of reported events revealed that in the past the reported scenario has never led to serious injury.During the performed investigation it was established that expected wear and tear occur naturally during normal operation and indicate that the squeeze tube is no longer functioning according to specification.According to the manual preventive maintenance on dosing system has to be performed yearly, including checking of detergent tubing, which should be replaced if it is required.Unfortunately, manufacturer could not conduct the technical investigation and determine the root cause without requested information.However, in case of new relevant information, the case will be reconsidered.We believe that devices in the market are performing correctly overall.Given the circumstances and the fact that there is no apparent trend in complaints of this nature, we shall continue to monitor the customer experiences with the devices for any future information.The purpose of this submission is also to provide a correction of h4.Device manufacture date this is based on the result of the review noting the initial report was incorrectly submitted stating another manufacture date.#h4 previous device manufacture date#: 2010-12-21.Corrected device manufacture date#: 2010-11-29.
 
Event Description
Manufacturer's reference number:(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
86-SERIES
Type of Device
DISINFECTOR, MEDICAL DEVICES
Manufacturer (Section D)
GETINGE DISINFECTION AB
ljungadalsgatan 11
vaxjo
Manufacturer (Section G)
GETINGE DISINFECTION AB
ljungadalsgatan 11
vaxjo
Manufacturer Contact
dennis genito
ljungadalsgatan 11
vaxjo 
MDR Report Key13059551
MDR Text Key282611515
Report Number9616031-2021-00040
Device Sequence Number1
Product Code MEC
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 12/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8666
Device Catalogue NumberS-8666913-CTOM
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received03/23/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/29/2010
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-