Model Number A1108
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
A facility reported that the swivel lock on the mayfield triad skull clamp (a1108) was able to rotate from the locked to unlocked position very easily and did not 'roll' into the locked position.It became unlocked while turning the patient.No patient injury or surgical delay has been reported.
 
Manufacturer Narrative
The mayfield triad skull clamp (a1108) was returned for evaluation: failure analysis - complaint confirmed via inspection of the unit.The index knob was loose and the unit required replacement of worn internal parts.Unit received with the lock having rotational and lateral movement and a residue buildup was present.New components have been added to replace all worn internal parts, and general maintenance and cleaning also performed.Root cause - complaint confirmed via inspection of the unit.The index knob was loose and the unit required replacement of worn internal parts.This unit is beyond integra¿s 7 years recommended life cycle (manufactured 2010).No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD TRIAD SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key13061313
MDR Text Key282634709
Report Number3004608878-2021-00693
Device Sequence Number1
Product Code HBL
UDI-Device Identifier10381780253518
UDI-Public10381780253518
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA1108
Device Catalogue NumberA1108
Device Lot Number109(2010)
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2021
Initial Date Manufacturer Received 11/29/2021
Initial Date FDA Received12/22/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-