• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH ENTRFLX 10FR;43IN W/STY YPORT; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH ENTRFLX 10FR;43IN W/STY YPORT; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 8884721088
Device Problem Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/27/2021
Event Type  malfunction  
Event Description
The customer reported that after insertion of the product by multiple highly skilled staff, the guidewire becomes stuck and is extremely difficult to remove even with flushing as recommended trouble shooting.The guide wire has a sticky feel as you withdraw it as if it is getting stuck on the inside of the tube.This has caused the tube to become dislodged and coil, requiring a new product second or even third re-insertion as the next product will have the same occurrence.Per additional information received, there was no patient injury or death reported.The patient required extra insertion due to the alleged issue.There was no change reported in the current status of the patient.Also, it was confirmed with clinical staff they are submerging the tube in water for 5 seconds prior to inserting it.
 
Manufacturer Narrative
The complainant indicated that the device will not be returned for evaluation; therefore, a failure analysis is not available, and we are not able to determine the relationship between this device and the cause for this event.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.If additional information or the sample is received, the investigation will be reopened and responded to accordingly.
 
Manufacturer Narrative
The lot number was provided, and the device history record (dhr) was reviewed indicating that the product was released accomplishing all quality standards.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.The product was manufactured on 01-jun-2021.A sample was not received for the investigation.Because a sample was not returned, we were unable to perform a follow up investigation to include functional and visual evaluations to confirm the issue and root cause analysis.During the investigation, a review through the manufacturing process was conducted.All process and controls were found properly followed, including sub-assemblies, finished product assembly and packaging.There were no abnormal conditions found that could trigger the reported condition.At this time, a corrective and preventive action is not deemed necessary.The current process is running according to product specifications, meeting all quality acceptance criteria.We will keep monitoring the process for any adverse trends that require immediate attention.This complaint will be used for qa tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTRFLX 10FR;43IN W/STY YPORT
Type of Device
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX  22500
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key13061944
MDR Text Key282634807
Report Number9612030-2021-03125
Device Sequence Number1
Product Code KNT
UDI-Device Identifier10884521518872
UDI-Public10884521518872
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8884721088
Device Catalogue Number8884721088
Device Lot Number2114502764
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/02/2021
Initial Date FDA Received12/22/2021
Supplement Dates Manufacturer Received12/02/2021
Supplement Dates FDA Received02/03/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2021
Is the Device Single Use? Yes
Type of Device Usage A
Patient Sequence Number1
-
-