MAUDE Adverse Event Report: INTEGRA - PRINCETON DURASEAL EXACT SPINE SEALANT SYSTEM 5ML US BOX OF 5; DURASEAL SPINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA - PRINCETON DURASEAL EXACT SPINE SEALANT SYSTEM 5ML US BOX OF 5; DURASEAL SPINE Back to Search Results
Catalog Number 206520
Device Problem Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/04/2021
Event Type  malfunction  
Event Description
A facility reported that when duraseal (206520) was used with the extended tip applicator, it worked great the first time.Then when it was handed back to spray again, the extended tip was clogged.Blue fluid could be seen on both sides of the applicator, indicating there is a fault within the extended tip.When a new kit was opened, it worked as it should with the first spray without extended tip.Then, after 2 minutes and a fresh regular tip, the kit was clogged.Clear fluid came out, as well as solidified blue which indicates that the y connector is faulty and is mixing in the y connector rather than the tip where it should be mixing.This is a recurring problem.There was a 15 minute delay without adverse patient outcome.
 
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
N/a.
 
Manufacturer Narrative
Duraseal was not returned for evaluation and lot number information has not been provided; therefore, an evaluation of the device could not be performed, and device history record (dhr) could not be reviewed.The root cause(s) of the reported issue could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURASEAL EXACT SPINE SEALANT SYSTEM 5ML US BOX OF 5
Type of Device
DURASEAL SPINE
Manufacturer (Section D)
INTEGRA - PRINCETON
1100 campus road
1100 campus road
princeton NJ
Manufacturer (Section G)
INTEGRA - PRINCETON
1100 campus road
none
princeton NJ
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key13061982
MDR Text Key284529581
Report Number3003418325-2021-00015
Device Sequence Number1
Product Code NQR
Combination Product (y/n)N
PMA/PMN Number
P080013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number206520
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/02/2021
Initial Date FDA Received12/22/2021
Supplement Dates Manufacturer Received02/23/2022
Supplement Dates FDA Received03/03/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient SexFemale
-
-