During internal review of programming data, a m1000 was found to have experienced an suspected gc5 reset.The reset was observed on (b)(6) 2021 @ 3:08 with reboot reason 6 and reboot counter of 2.This event is consistent with the issue associated with the gc5 recall which affects some m1000 generators.For these events, the root cause of the unexpected device reboot is related to a hardware bug within the microcontrollers used in these devices and interaction with the device firmware.The device history records of the generator were reviewed.The generator passed final quality and functional specifications prior to release.The generator was manufactured before the screening process was implemented to mitigate the potential for gc5 reboots.No additional relevant information has been received.
|