PERFUSION SYSTEMS CARDIOBLATE; SURGICAL DEVICE, FOR CUTTING, COAGULATION, AND/OR ABLATION OF TISSUE, INCLUDING
|
Back to Search Results |
|
Model Number 49260 |
Device Problem
Material Separation (1562)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/01/2021 |
Event Type
malfunction
|
Event Description
|
Medtronic received information that during use of this cardioblate gemini probe when the customer was on the third clamp, the end of the probe broke.The second half (broken part) remained on the guide, this guide was removed along with it.The probe was then removed and replaced in order to complete the procedure.There were no adverse patient effects as a result of this issue.Customer stated the patient's large chest and large heart cavities may have been the cause, as the patient was very large.
|
|
Manufacturer Narrative
|
Device evaluation summary: visual inspection shows the device was returned without the broken off piece of the tip and the handle is apart with a loose piece.Additional visual inspection of the remining jaw shows little to no evidence of any damage/tool marks.Reason for return was confirmed.Investigation conclusion: product analysis showed that potential causes of the damage may be due to user issues: use of forceps, manipulated until it fractured (the tips are not malleable like the bp2 device, so they are not able to be reshaped), excessive force placed on the tip or the tip was exposed to severe shock during use and/or storage and handling, or potentially clasped onto something hard other than tissue.The ifu has the following warning; ¿the device may be damaged before or during the procedure by improper handling or other intervening acts.¿ there were no patient adverse effects, will continue to monitor for future events.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Further investigation determined, an investigation has been opened into the issue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank, because the information is currently, unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|