• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS LUCERA DUODENOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS LUCERA DUODENOVIDEOSCOPE Back to Search Results
Model Number TJF-260V
Device Problems Device Reprocessing Problem (1091); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject device was inspected at olympus service operation repair center (sorc).Sorc checked the subject device and found that the reported phenomenon was duplicated.The channel cleaning brush could not be inserted smoothly into the instrument channel due to the clogging of the instrument channel.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.However, based on the reported information, there was possibility that the foreign material that entered the instrument channel from the outside could not be sufficiently removed due to insufficient cleaning and rinsing.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed from the user that during the unspecified timing, the foreign matter came out from the channel.The user stated that stent had remained in the instrument channel outlet.There was possibility that insufficiently or incorrectly reprocessed subject device might be used for next procedure.The occurrence date of the event is unknown.There was no report of patient injury associated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13086834
MDR Text Key282847082
Report Number8010047-2021-16876
Device Sequence Number1
Product Code FDT
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTJF-260V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2021
Was the Report Sent to FDA? No
Date Manufacturer Received12/03/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/19/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-