Model Number 26605
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/08/2021
Event Type  malfunction  
Event Description
Reportable based on device analysis completed on 08dec2021.It was reported that partial deployment occurred.The target lesion was located in a carotid artery.A 10.0-24 carotid wallstent was selected for used.However, during unpacking, it was noticed that stent was partially deployed.The procedure was completed with another of same device.No patient complications were reported.However, device analysis revealed that stent fully deployed and detached from the delivery system.
 
Manufacturer Narrative
Device evaluated by mfr.: a carotid device was returned for analysis.A visual and tactile examination identified no issues with the catheter or delivery system that could potentially have contributed to the complaint incident.The device was returned with the stent fully deployed and detached from the delivery system.No damage or issues were noted with the deployed stent.This concludes the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAROTID WALLSTENT
Type of Device
STENT, CAROTID
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13087263
MDR Text Key282781937
Report Number2134265-2021-16189
Device Sequence Number1
Product Code NIM
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 12/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number26605
Device Catalogue Number26605
Device Lot Number0027039484
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/25/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/08/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/26/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age58 YR
Patient SexFemale
Patient Weight65 KG
-
-