During the atrial flutter (afl) ablation procedure, the tactisys was connected to the dws and when the catheter was connected to the tactisys.Towards the end of the procedure, saline began to leak into the light pipeline and the electrical signal had interference.The device was replaced and the procedure was completed with no adverse consequences to the patient.
|
One 75 mm tacticath quartz contact force ablation catheter was received for evaluation.Fluid was noted outside of the irrigation tubing, the device did not meet specifications during an irrigation leak test.In addition, electrodes 1-4 met specifications during electrical testing, however, multiple short circuits were detected.Dried saline was noted within the 19-pin connector, consistent with the observed short circuits.Further testing revealed a leak at the luer/irrigation tubing junction, consistent with the fluid noted in the proximal tubing, fluid ingress into the 19-pin connector, and the reported leak.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the leak at the luer/irrigation tubing junction remains unknown.
|