ZIMMER SURGICAL, INC. FAN SPRAY KIT; LAVAGE, JET Back to Search Results
Model Number N/A
Device Problems Inability to Irrigate (1337); Battery Problem (2885); Expulsion (2933)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/20/2021
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Functional testing determined that the device would not power on when connected to the original battery pack, but did power on and function normally when connected to the lab battery pack.Visual examination of the returned product identified that 2 batteries from the battery pack had leaked electrolyte onto the terminals inside the pack.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.Device is used for treatment.Anode expulsions likely due to a short in the electrical wiring, but a definitive root cause cannot be determined.The event is confirmed.
 
Event Description
It was reported that the new fan spray kit didn't work when the surgeon turned on the switch.Investigation found two batteries from the battery pack had leaked electrolyte onto the terminals inside the pack.No adverse event was reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FAN SPRAY KIT
Type of Device
LAVAGE, JET
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key13090180
MDR Text Key282908441
Report Number0001526350-2021-01390
Device Sequence Number1
Product Code FQH
UDI-Device Identifier00889024375161
UDI-Public(01)00889024375161(17)231022(10)64911827
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial
Report Date 12/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/22/2023
Device Model NumberN/A
Device Catalogue Number00515047501
Device Lot Number64911827
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/08/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/22/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/22/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-