B. BRAUN MEDICAL INC. SAFEDAY¿; SET, ADMINISTRATION, INTRA Back to Search Results
Model Number 352894
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event Description
As reported by the user facility: it was reported that the set has a dirty chamber.There is brown rust colored dirt inside of the chamber.No injury reported.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).One (1) used sample without packaging was provided for evaluation.A visual evaluation was performed and under magnification, the material was found to be embedded in the molding of the drip chamber.The supplier was contacted and based on the suppliers evaluation the "foreign material" was confirmed to be embedded degraded pvc( material of the soft chamber) and would not would not affect the functionality of the part and would not have contribute to contamination of drip chamber.A review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.In addition, review of the batch history records was also performed for the reported lot number and no non-conformances or deviations were noted during the manufacturing process or final inspections.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFEDAY¿
Type of Device
SET, ADMINISTRATION, INTRA
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
Manufacturer (Section G)
B. BRAUN DOMINICIAN REPULIC INC.
las americas industrial park
km22 autopista las americas
santo domingo
DR  
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key13090734
MDR Text Key283317125
Report Number2523676-2021-00416
Device Sequence Number1
Product Code FPA
UDI-Device Identifier04046964181894
UDI-Public(01)04046964181894
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K955585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Model Number352894
Device Catalogue Number352894
Device Lot Number0061739315
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/15/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-