This device is not distributed in us so that 510k# is blank.We checked the returned unit and confirmed the ccd module as defective.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module.In addition, we confirmed that the bending rubber leakage, the insertion flexible tube (ift) buckled, the root brace rubber cut, and the remote control buttons cut; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.
|