Model Number E100-25M
Device Problem Material Split, Cut or Torn (4008)
Patient Problem Mitral Valve Insufficiency/ Regurgitation (4451)
Event Date 12/02/2021
Event Type  Injury  
Event Description
It was reported that in 2016, a 25mm epic stented porcine heart valve was implanted as part of a double valve replacement.On (b)(6) 2021, "the patient was raced to the hospital due to cardiac insufficiency.The patient underwent a semi-emergent surgery for redo mvr." the epic valve was explanted.Upon explant, the leaflet consistent with p3 position had become torn at the base of the leaflet and free wall (tip).No indication of infection was observed on the valve.A non-abbott valve was successfully implanted.The patient status was reported as unknown.No additional information is available.
 
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
 
Manufacturer Narrative
An event of cardiac insufficiency, torn cusp, and valve explant was reported.A more comprehensive assessment could not be performed as the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC STENTED PORCINE HEART VALVE W/FLEXFIT SYSTEM
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ST. JUDE MEDICAL, BRASIL LTDA.
rua professor jose vieira de mendonça 1301
engenho nogueira - belo horizonte - mg
belo horizonte 31310 -260
BR  31310-260
Manufacturer (Section G)
ST. JUDE MEDICAL, BRASIL LTDA.
rua professor jose vieira de mendonça 1301
engenho nogueira - belo horizonte - mg
belo horizonte 31310 -260
BR   31310-260
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key13095080
MDR Text Key282825435
Report Number3001883144-2021-00150
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P040021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberE100-25M
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/30/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
-
-