ABBOTT QUADRA ASSURA MP ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3371-40QC
Device Problems Failure to Capture (1081); Over-Sensing (1438); No Pacing (3268)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/13/2021
Event Type  malfunction  
Event Description
Related manufacturer report numbers: 2017865-2021-40173, 2017865-2021-40174.During an in clinic follow-up, loss of capture resulting in pacing inhibition and episodes of oversensing was observed on the right ventricular (rv) lead, left ventricular (lv) lead and the device.Abbott technical support was contacted and the device was reprogrammed to resolve the event.The patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key13097596
MDR Text Key283089315
Report Number2017865-2021-40175
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodePO
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2018
Device Model NumberCD3371-40QC
Device Lot NumberP000025570
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/13/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/09/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-