Model Number CD3371-40QC
Device Problem Over-Sensing (1438)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the patient presented remotely via merlin.Net.Upon review of the transmission, it was observed that the implantable cardioverter-defibrillator (icd) exhibited non-sustained myopotential oversensing which led to inhibition of pacing.The patient did not receive inappropriate therapy.No programming changes were made.There were no patient consequences.
 
Event Description
New information noted that the clinic called asking about the episodes of rvos.The decision was made to continue to monitor the lead noise as the patient did not experience any pacing inhibition at this time.
 
Event Description
New information noted that the programming interventions were made.The patient was in stable condition before, during, and after the interrogation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key13098232
MDR Text Key283089235
Report Number2017865-2021-40002
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date 01/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2021
Device Model NumberCD3371-40QC
Device Lot NumberA000073901
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/21/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/07/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
OPTISURE LEAD
-
-