MAUDE Adverse Event Report: OAKDALE LEVEL 1 HOTLINE 390 FLUID WARMING DEVICE; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 HOTLINE 390 FLUID WARMING DEVICE; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-390
Device Problem No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported the device lcd was not working.No patient was involved.
 
Manufacturer Narrative
Other, other text: the failure resulted from physical damage sustained by the device during use and handling.The device history record (dhr) review is not applicable or relevant because the results of the complaint investigation do not indicate a problem with the initial manufacture or prior repair of the device.A product sample was received for evaluation.Visual and functional testing were performed.Visual inspection found a noisy pump, damaged printed circuit board (pcb), dirty enclosure, cracked tank cover, rusty heater, bent micro switch, broken front cover, and a faded line cord.Functional testing found the display was blank after powering on.The root cause of the reported issue was found to be customer damage to the lcd and printed circuit board (pcb) from impact during use and handling.Replaced the front cover, line cord, micro switch, heater, tank cover, enclosure, printed circuit board (pcb), and pump.The device passed all functional testing.
 
Event Description
Additional information was received which included that the event date was unknown and there was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE 390 FLUID WARMING DEVICE
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
7633833310
MDR Report Key13101435
MDR Text Key282899921
Report Number3012307300-2021-13364
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier10695085002796
UDI-Public10695085002796
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-390
Device Catalogue NumberCON-HL-390
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2021
Was the Report Sent to FDA? No
Date Manufacturer Received07/01/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-