Other, other text: the failure resulted from physical damage sustained by the device during use and handling.The device history record (dhr) review is not applicable or relevant because the results of the complaint investigation do not indicate a problem with the initial manufacture or prior repair of the device.A product sample was received for evaluation.Visual and functional testing were performed.Visual inspection found a noisy pump, damaged printed circuit board (pcb), dirty enclosure, cracked tank cover, rusty heater, bent micro switch, broken front cover, and a faded line cord.Functional testing found the display was blank after powering on.The root cause of the reported issue was found to be customer damage to the lcd and printed circuit board (pcb) from impact during use and handling.Replaced the front cover, line cord, micro switch, heater, tank cover, enclosure, printed circuit board (pcb), and pump.The device passed all functional testing.
|