DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,INTL; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-UC-3013-55 |
Device Problem
Failure to Zero (1683)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/07/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate this unit and found that the pressure-trunk cable was defective and need to be replaced.In addition, the pressure cable is not a getinge part, and is from another vendor.Once the customer purchased it, they will revert to us.Repairs are ongoing and a supplemental report will be submitted when additional information is provided.
|
|
Event Description
|
It was reported that during routine check, the cs100 intra-aortic balloon pump (iabp) was not able to zero pressure.There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Updated fields: b4, g3, g6, g7, h2, h4, h6 (type of investigation, investigation findings, investigation conclusions), h10, h11.Corrected fields: d1, d4, d5, h6 (component codes).A getinge field service engineer (fse) was dispatched to evaluate the iabp unit and was able to reproduce the reported issue.The fse found a problem with pressure cable which is defective which has to be replaced.Pressure cable is not a getinge part, and it is from some other vendor.Once customer purchases it, they will revert to them.The fse conversed over the phone with the customer who informed them, the customer has purchased required pressure cable and the iabp unit is now functional.
|
|
Event Description
|
N/a.
|
|