• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,INTL; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS100 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,INTL; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-UC-3013-55
Device Problem Failure to Zero (1683)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2021
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to evaluate this unit and found that the pressure-trunk cable was defective and need to be replaced.In addition, the pressure cable is not a getinge part, and is from another vendor.Once the customer purchased it, they will revert to us.Repairs are ongoing and a supplemental report will be submitted when additional information is provided.
 
Event Description
It was reported that during routine check, the cs100 intra-aortic balloon pump (iabp) was not able to zero pressure.There was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
Updated fields: b4, g3, g6, g7, h2, h4, h6 (type of investigation, investigation findings, investigation conclusions), h10, h11.Corrected fields: d1, d4, d5, h6 (component codes).A getinge field service engineer (fse) was dispatched to evaluate the iabp unit and was able to reproduce the reported issue.The fse found a problem with pressure cable which is defective which has to be replaced.Pressure cable is not a getinge part, and it is from some other vendor.Once customer purchases it, they will revert to them.The fse conversed over the phone with the customer who informed them, the customer has purchased required pressure cable and the iabp unit is now functional.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS100 INTRA-AORTIC BALLOON PUMP, REFURBISHED, ENGLISH,NON-UTS,INTL
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key13102151
MDR Text Key284980379
Report Number2249723-2021-02977
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K031636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 06/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-UC-3013-55
Device Catalogue Number0998-UC-3013-55
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/27/2005
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-