ABBOTT QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problems Break (1069); Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/17/2021
Event Type  malfunction  
Event Description
It was reported that during the initial implant procedure, the stylet was unable to be removed from the left ventricular (lv) lead resulting in lead damage while attempting to the remove the stylet.The lead was explanted and replaced to resolve the event.The patient was in stable condition.
 
Manufacturer Narrative
The reported events of stylet was unable to be removed and lead damage while attempting to the remove the stylet were confirmed.As received, a complete lead was returned in one piece with the stylet stuck inside the lead.The cause of the reported event of lead damage while attempting to the remove the stylet was isolated to procedural damage.The cause of the reported event of stylet was unable to be removed was isolated to the bunching of stripped stylet, ptfe coating and inner coil.A review of the device history record (dhr) confirmed that no issues were identified related to this reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key13108998
MDR Text Key282910179
Report Number2017865-2021-40286
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeGM
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberS000084679
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/04/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/11/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-