Model Number CD3365-40Q
Device Problem Defective Device (2588)
Patient Problem Insufficient Information (4580)
Event Date 12/17/2021
Event Type  malfunction  
Event Description
This report is to advise of an event observed during analysis.
 
Manufacturer Narrative
No complaint was received with the return of the device.Failure event observed during analysis.Final analysis found that as received, the device was returned to the lab.Longevity calculation found that battery depletion was premature based on device usage.Electrical testing revealed the cause of the event was high current drain from the hybrid.Review of device history records did not identify any non-conformances related to the event.No other anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA CRT-D QUAD RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key13111981
MDR Text Key282919962
Report Number2017865-2021-39969
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508292
UDI-Public05414734508292
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date06/30/2018
Device Model NumberCD3365-40Q
Device Catalogue NumberCD3365-40Q
Device Lot NumberA000024046
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/26/2021
Was the Report Sent to FDA? No
Date Manufacturer Received12/17/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/11/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-