ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CDDRA500Q
Device Problems Environmental Compatibility Problem (2929); Inappropriate or Unexpected Reset (2959)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/09/2021
Event Type  malfunction  
Event Description
It was reported that the patient underwent a magnetic resonance imaging (mri) scan.The implantable cardioverter defibrillator was not programmed into mri mode and the device went into backup mode.The high voltage therapy was disabled as a result.The device was reset.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of bvvi following mri was confirmed.The device was not programmed to mri settings.Based on analysis of the device image information received, the device entered power on reset due to the mri settings not being enabled by the user prior to mri.Additionally, there is no evidence of preventative device programming prior to the mri scan as tachy therapy and magnet response remained enabled.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT DR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key13113827
MDR Text Key283147618
Report Number2017865-2021-40284
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032003
UDI-Public05415067032003
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Model NumberCDDRA500Q
Device Catalogue NumberCDDRA500Q
Device Lot NumberP000123658
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/04/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/28/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
OPTISURE; TENDRIL
Patient Age68 YR
Patient SexFemale
-
-