• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 PUREWICK URINE COLLECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 PUREWICK URINE COLLECTION SYSTEM Back to Search Results
Model Number PW200
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/16/2021
Event Type  malfunction  
Event Description
It was reported that the purewick urine collection system motor was hot and smelled hot.It was noted that the patient had been using the purewick products for less than 30 days.
 
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Manufacturer Narrative
The reported event was inconclusive.No sample was returned for evaluation.A potential root cause for this failure could be " inadequate component selection component failures, damage, or deterioration ".The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: " always unplug purewick¿ urine collection system before cleaning or when not in use.Do not immerse the purewick¿ urine collection system in water.As with most electrical devices, electrical parts in this system are electrically live even when the power is off.To reduce the risk of electric shock, if the purewick¿ urine collection system falls into water, unplug immediately.Do not reach into the water to retrieve it." " warning: this device should not be used in oxygen rich environments or in conjunction with flammable anesthetics." " when the purewick¿ urine collection system has been stored in a very cold environment (down to -13°f or -25°c), please allow the device to sit for approximately 90 minutes at room temperature (approximately 68°f or 20°c) before use." " when the purewick¿ urine collection system has been stored in a very warm environment (up to 158°f or 70°c), please allow the device to sit for approximately 90 minutes at room temperature (approximately 68°f or 20°c) before use.¿ h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event Description
It was reported that the purewick urine collection system motor was hot and smelled hot.It was noted that the patient had been using the purewick products for less than 30 days.Per follow up information via phone on 28mar2022, customer stated that everything has been resolved.Troubleshooting helped to determine the smell was not from the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUREWICK URINE COLLECTION SYSTEM
Type of Device
PUREWICK URINE COLLECTION SYSTEM
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key13114420
MDR Text Key284526956
Report Number1018233-2021-08526
Device Sequence Number1
Product Code NZU
UDI-Device Identifier00801741185359
UDI-Public(01)00801741185359
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberPW200
Device Catalogue NumberPW200
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/16/2021
Initial Date FDA Received12/29/2021
Supplement Dates Manufacturer Received03/25/2022
Supplement Dates FDA Received04/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
-
-