MAUDE Adverse Event Report: MICROPORT CRM S.R.L. PLATINIUM CRT-D 1711; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER,WITH CRT-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROPORT CRM S.R.L. PLATINIUM CRT-D 1711; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER,WITH CRT-D Back to Search Results
Model Number PLATINIUM CRT-D 1711
Device Problem Over-Sensing (1438)
Patient Problems Unspecified Heart Problem (4454); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/24/2020
Event Type  malfunction  
Event Description
Reportedly, ventricular noise oversensing was suspected due to myopotential interactions.
 
Manufacturer Narrative
Analysis of the provided patient files confirmed that some episodes recorded in the device memory showed a ventricular noise oversensing resulting from electromagnetic interferences and/or myopotentials, were observed in the majority of the episodes recorded in the device since (b)(6) 2020.Based on the available data, no anomaly is suspected on the crt-d.However, careful monitoring of this phenomenon should be performed upon each follow-up.It should be noted that reprogramming of vt/vf detection parameters (such as extending the persistence) should be weighed against the risk of delaying appropriate therapy.
 
Event Description
For a platinium, ventricular oversensing due to myopotential interactions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLATINIUM CRT-D 1711
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER,WITH CRT-D
Manufacturer (Section D)
MICROPORT CRM S.R.L.
via crescentino s.n
saluggia (vc) 13040
IT  13040
Manufacturer (Section G)
MICROPORT CRM S.R.L.
via crescentino s.n
saluggia (vc) 13040
IT   13040
Manufacturer Contact
elodie vincent
via crescentino s.n
saluggia (vc) 13040
IT   13040
MDR Report Key13117683
MDR Text Key283080274
Report Number1000165971-2021-01580
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P060027
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPLATINIUM CRT-D 1711
Device Catalogue NumberPLATINIUM CRT-D 1711
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/03/2021
Initial Date FDA Received12/30/2021
Supplement Dates Manufacturer Received01/03/2022
Supplement Dates FDA Received01/10/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient EthnicityNon Hispanic
-
-