MICROPORT CRM S.R.L. PLATINIUM CRT-D 1711; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER,WITH CRT-D
|
Back to Search Results |
|
Model Number PLATINIUM CRT-D 1711 |
Device Problem
Over-Sensing (1438)
|
Patient Problems
Unspecified Heart Problem (4454); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/24/2020 |
Event Type
malfunction
|
Event Description
|
Reportedly, ventricular noise oversensing was suspected due to myopotential interactions.
|
|
Manufacturer Narrative
|
Analysis of the provided patient files confirmed that some episodes recorded in the device memory showed a ventricular noise oversensing resulting from electromagnetic interferences and/or myopotentials, were observed in the majority of the episodes recorded in the device since (b)(6) 2020.Based on the available data, no anomaly is suspected on the crt-d.However, careful monitoring of this phenomenon should be performed upon each follow-up.It should be noted that reprogramming of vt/vf detection parameters (such as extending the persistence) should be weighed against the risk of delaying appropriate therapy.
|
|
Event Description
|
For a platinium, ventricular oversensing due to myopotential interactions.
|
|