• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC LIBRE 2 SENSOR FREESTYLE; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC LIBRE 2 SENSOR FREESTYLE; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/27/2021
Event Type  malfunction  
Manufacturer Narrative
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A customer reported receiving erroneous glucose results from an abbott diabetes care device.The results when plotted on a parkes error grid fell into either the c, d, or e zone.There was no report of death, serious injury, or mistreatment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBRE 2 SENSOR FREESTYLE
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key13117929
MDR Text Key282966304
Report Number2954323-2021-97390
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? No
Date Manufacturer Received12/27/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/11/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Patient Sequence Number1
-
-