The mayfield skull clamp (a1059) was returned for evaluation: device history record (dhr) - the dhr was reviewed and no anomalies related to the reported failure was observed.Failure analysis - the investigation of the returned device showed that the reported complaint was confirmed from the evaluation.Unit received with the lock having both rotational and lateral movement and a residue buildup was present.Upon disassembly, repair noted the index knob, and the lock will need new components added to replace worn internal parts.Unit was machined to have heli-coils added to the large starburst threads.New components added to replace worn internal parts along with general maintenance and cleaning performed.Root cause - complaint confirmed via inspection of the item.The index knob was unable to be rotated into the fully locked position and could only be partially rotated, meaning the skull clamp could not be fully locked.Additionally, unit does not have a service history and it is recommended to have service performed.No further investigation required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.
|