• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Model Number A1059
Device Problem Device Slipped (1584)
Patient Problem Laceration(s) (1946)
Event Date 12/10/2021
Event Type  Injury  
Event Description
A facility reported that the mayfield skull clamp (a1059) slipped and caused a patient laceration.No further information has been provided.
 
Manufacturer Narrative
The mayfield skull clamp (a1059) was returned for evaluation: device history record (dhr) - the dhr was reviewed and no anomalies related to the reported failure was observed.Failure analysis - the investigation of the returned device showed that the reported complaint was confirmed from the evaluation.Unit received with the lock having both rotational and lateral movement and a residue buildup was present.Upon disassembly, repair noted the index knob, and the lock will need new components added to replace worn internal parts.Unit was machined to have heli-coils added to the large starburst threads.New components added to replace worn internal parts along with general maintenance and cleaning performed.Root cause - complaint confirmed via inspection of the item.The index knob was unable to be rotated into the fully locked position and could only be partially rotated, meaning the skull clamp could not be fully locked.Additionally, unit does not have a service history and it is recommended to have service performed.No further investigation required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key13119211
MDR Text Key285613541
Report Number3004608878-2021-00711
Device Sequence Number1
Product Code HBL
UDI-Device Identifier10381780253457
UDI-Public10381780253457
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberA1059
Device Catalogue NumberA1059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2021
Date Manufacturer Received12/10/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/08/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-