• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INOGEN X4 CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INOGEN X4 CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR Back to Search Results
Model Number G148
Device Problems Premature Discharge of Battery (1057); Defective Device (2588); Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/25/2021
Event Type  Injury  
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) was exhibiting beeping.In addition, the patient reported device heating, which technical service described as atypical behavior.During medical consult, it was determined that this device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.Data analysis showed the battery appeared to be depleting more quickly than expected, and a boston scientific technical services (ts) consultant recommended device replacement.This crt-d was explanted and successfully replaced.No additional adverse patient effects were reported.
 
Manufacturer Narrative
Quality assurance laboratory.Review of the device memory confirmed that a low voltage alert was recorded.Although the device memory diagnostics demonstrated that the daily battery voltage measurements displayed an irregular pattern of discharge, the battery voltage level upon receipt in the laboratory remained sufficient to ensure therapy availability/delivery while the device was implanted.Using historical daily battery voltage measurement data, engineers determined that this device was demonstrating behavior consistent with a latent current leakage path within the battery itself, which resulted in a partial depletion of the battery.This behavior is caused by a latent electrical leakage path that develops over time between the anode and cathode within the device battery.In certain circumstances, lithium metal ions can re-plate to form clusters that may result in an internal current leakage.A design enhancement was implemented in 2013 to enlarge the physical barrier between the electrode layers to prevent current leakage in the area of the battery that is most often affected by this behavior.
 
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) was exhibiting beeping.In addition, the patient reported device heating, which technical service described as atypical behavior.During medical consult, it was determined that this device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.Data analysis showed the battery appeared to be depleting more quickly than expected, and a boston scientific technical services (ts) consultant recommended device replacement.This crt-d was explanted and successfully replaced.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN X4 CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key13137951
MDR Text Key283180818
Report Number2124215-2021-36515
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534591
UDI-Public00802526534591
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/07/2018
Device Model NumberG148
Device Catalogue NumberG148
Device Lot Number143913
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/01/2021
Initial Date FDA Received01/01/2022
Supplement Dates Manufacturer Received10/25/2022
Supplement Dates FDA Received10/27/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2016
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age61 YR
Patient SexMale
-
-