Model Number G148 |
Device Problems
Premature Discharge of Battery (1057); Defective Device (2588); Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/25/2021 |
Event Type
Injury
|
Event Description
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) was exhibiting beeping.In addition, the patient reported device heating, which technical service described as atypical behavior.During medical consult, it was determined that this device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.Data analysis showed the battery appeared to be depleting more quickly than expected, and a boston scientific technical services (ts) consultant recommended device replacement.This crt-d was explanted and successfully replaced.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Quality assurance laboratory.Review of the device memory confirmed that a low voltage alert was recorded.Although the device memory diagnostics demonstrated that the daily battery voltage measurements displayed an irregular pattern of discharge, the battery voltage level upon receipt in the laboratory remained sufficient to ensure therapy availability/delivery while the device was implanted.Using historical daily battery voltage measurement data, engineers determined that this device was demonstrating behavior consistent with a latent current leakage path within the battery itself, which resulted in a partial depletion of the battery.This behavior is caused by a latent electrical leakage path that develops over time between the anode and cathode within the device battery.In certain circumstances, lithium metal ions can re-plate to form clusters that may result in an internal current leakage.A design enhancement was implemented in 2013 to enlarge the physical barrier between the electrode layers to prevent current leakage in the area of the battery that is most often affected by this behavior.
|
|
Event Description
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) was exhibiting beeping.In addition, the patient reported device heating, which technical service described as atypical behavior.During medical consult, it was determined that this device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.Data analysis showed the battery appeared to be depleting more quickly than expected, and a boston scientific technical services (ts) consultant recommended device replacement.This crt-d was explanted and successfully replaced.No additional adverse patient effects were reported.
|
|
Search Alerts/Recalls
|
|