Catalog Number 80202
Device Problems Break (1069); Failure to Advance (2524); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2021
Event Type  malfunction  
Manufacturer Narrative
The catalog number identified has not been cleared in the us but is similar to the rotarex that are cleared in the us.The pro code and 510 k number for the rotarex are identified in sections procode and pma/510k.As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, photo and image were provided for review.The investigation of the reported event is currently underway.The medical device manufacturer and manufacturing location for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.
 
Event Description
It was reported that during a recanalization procedure through the right groin to treat stent stenosis of the left sfa region all the way to the popliteal, the device allegedly faced resistance when advancing.The device was pulled back when it was noted that the device got stuck with the hemostatic valve and the helix inside the catheter was noted to be broken.The procedure was completed using another device.There was no reported patient injury.
 
Event Description
It was reported that during a recanalization procedure through the right groin to treat stent stenosis of the left sfa region all the way to the popliteal, the device allegedly faced resistance when advancing.The device was pulled back when it was noted that the device got stuck with the hemostatic valve and the helix inside the catheter was noted to be broken.The procedure was completed using another device.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the rotarex that are cleared in the us.The pro code and 510 k number for the rotarex are identified in d2 and g4.H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was returned for evaluation.A physical investigation was performed for the catheter.The helix was broken at 28 cm distal.It can be confirmed that the helix was broken.Therefore, the investigation is confirmed for break and inconclusive for failure to advance and physical resistance/sticking.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: d4 (expiry date: 12/2023).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAREX
Type of Device
THROMBECTOMY & ATHERECTOMY
Manufacturer (Section D)
UNKNOWN
BR 
Manufacturer (Section G)
UNKNOWN
BR  
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key13148445
MDR Text Key285467587
Report Number3008439199-2021-00212
Device Sequence Number1
Product Code MCW
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K172315
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/03/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number80202
Device Lot Number210266
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/14/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-