BOSTON SCIENTIFIC CORPORATION PROMUS PREMIER SELECT; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number 10666
Device Problems Positioning Failure (1158); Failure to Advance (2524); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/29/2021
Event Type  malfunction  
Event Description
It was reported that the struts were damaged.A 28 x 4.00 promus premier select drug-eluting stent was selected for use in a percutaneous transluminal coronary angioplasty (ptca).During the procedure, the stent was advanced, but could not pass through the lesion properly.The device was positioned at the lesion site, but the stent failed to deploy.The physician removed the stent from patient and noticed the struts were damaged.The procedure completed successfully with another same device.There were no patient complications reported.
 
Event Description
It was reported that the struts were damaged.A 28 x 4.00 promus premier select drug-eluting stent was selected for use in a percutaneous transluminal coronary angioplasty (ptca).During the procedure, the stent was advanced but could not pass through the lesion properly.The device was positioned at the lesion site but the stent failed to deploy.The physician removed the stent from patient and noticed the struts were damaged.The procedure completed successfully with another same device.There were no patient complications reported.It was further reported that the 55% stenosed target lesion was located in mildly tortuous and none calcified lesion.There were a multiple attempts made to position the stent at the lesion site.There was no any attempt made to inflate the balloon.
 
Manufacturer Narrative
Device evaluated by mfr: a promus premier select 28 x 4.00mm stent delivery system (sds) returned for analysis.A visual examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal markerbands.The stent od (outer diameter) was measured and the result was within max crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of distal tip damage.A visual and tactile examination of the hypotube shaft found no issues.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.
 
Event Description
It was reported that the struts were damaged.A 28 x 4.00 promus premier select drug-eluting stent was selected for use in a percutaneous transluminal coronary angioplasty (ptca).During the procedure the stent was advanced but could not pass through the lesion properly.The device was positioned at the lesion site, but the stent failed to deploy.The physician removed the stent from patient and noticed the struts were damaged.The procedure completed successfully with another same device.There were no patient complications reported.It was further reported that the 55% stenosed target lesion was located in mildly tortuous and none calcified lesion.There were a multiple attempts made to position the stent at the lesion site.There was no any attempt made to inflate the balloon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS PREMIER SELECT
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13159238
MDR Text Key283253427
Report Number2134265-2022-00006
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeTU
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 02/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Model Number10666
Device Catalogue Number10666
Device Lot Number0027449198
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/30/2021
Initial Date FDA Received01/04/2022
Supplement Dates Manufacturer Received01/05/2022
01/26/2022
Supplement Dates FDA Received01/19/2022
02/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/31/2021
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-