Catalog Number 1120250-23 |
Device Problems
Break (1069); Failure to Advance (2524); Deformation Due to Compressive Stress (2889)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/10/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.The additional xience alpine is filed under a separate medwatch report number.
|
|
Event Description
|
It was reported that the procedure was to treat a heavily calcified, heavily tortuous left anterior descending artery.Two xience alpine stent delivery systems (2.5x23mm and 2.75x33mm) failed to cross and both shafts on each device were noted to be "fractional" (break) and kinked.Another unspecified stent was used to successfully complete the procedure.There was no adverse patient effects reported and no clinically significant delay reported.No additional information was provided.
|
|
Manufacturer Narrative
|
A visual and dimensional inspection was performed on the returned device.The reported shaft break was not confirmed.The reported deformation due to compressive stress was confirmed.The reported failure to advance could not be tested as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation determined the reported failure to advance and deformation due to compressive stress appear to be related to circumstances of the procedure.The investigation was unable to determine a conclusive cause for the reported shaft break.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
|
Search Alerts/Recalls
|