SYNCARDIA SYSTEMS, LLC SYNCARDIA 70CC TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T); BIVENTRICULAR REPLACEMENT DEVICE
|
Back to Search Results |
|
Catalog Number 500101 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Hemolysis (1886); Hemorrhage/Bleeding (1888); Unspecified Infection (1930); Thrombosis/Thrombus (4440); Renal Impairment (4499); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Type
Injury
|
Manufacturer Narrative
|
The syncardia 70cc temporary total artificial heart (tah-t) is an implantable pulsatile biventricular replacement device that replaces a patient's native ventricles and valves and pumps blood to both the pulmonary and systemic circulation systems.The syncardia 70cc tah-t is indicated for use as a bridge to transplantation in cardiac transplant-eligible candidates at risk of imminent death from biventricular failure.The syncardia tah-t system is intended for use inside and outside the hospital.Investigation summary: intermacs patient registry data collected from april 1, 2021 through june 30, 2021 regarding adverse events was reviewed.Patients were de-identified therefore a one-to-one correlation could not be made between the patient and the lot number of the implanted tah-t.With a review of the available information, there is no evidence of a device malfunction or performance issues that would impact the reported events.Possible clinical factors that may have contributed to these events include the patients pre-existing history and related comorbidities (see section b7), the progression of their underlying disease, issues related to the therapeutic use of anticoagulants, antiplatelet medications and the patient's complex post-operative course.There are possible patient, pharmacological and procedural factors that may have contributed to these events.This device is used for treatment, not diagnosis.Ce (b)(4) pt (b)(6) initial.
|
|
Event Description
|
While supported by the tah-t, the patient experienced the following adverse events as defined by intermacs: 4 days post implant - renal dysfunction.6 days post implant - bleeding.9 days post implant - other sae.As of (b)(6) 2021 it was reported that the patient was still supported by the tah-t.
|
|
Event Description
|
During the last reporting period, it was reported that the patient had experienced the following adverse events as defined by intermacs.6 days post implant - hemolysis, 6 days post implant - right heart failure, 28 days post implant - hemolysis, 90 days post implant - hemolysis, 92 days post implant - device malfunction and/or pump thrombosis, 157 days post implant - infection - unk, 181 days post implant - hemolysis.The patient subsequently received a heart transplant after 214 days of tah-t support.
|
|
Manufacturer Narrative
|
Intermacs patient registry data collected from (b)(6) 2021 through (b)(6) 2022 regarding adverse events was reviewed.Patients were de-identified therefore a one-to-one correlation could not be made between the patient and the lot number of the implanted tah-t.Ra-1788, pt (b)(6), (di_468) follow-up report 1.
|
|
Search Alerts/Recalls
|
|
|