Catalog Number 227214
Device Problems Break (1069); Blocked Connection (2888); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/24/2021
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).Investigation summary: the complaint device was received at the service center and evaluated.It was reported that the staff was attempting to set up vapr wireless footpedal prior to case and footpedal did not pair.Changed batteries and tried to repair device, but still receiving error message.Per service reports, this complaint can be confirmed.During the service evaluation the following defects were identified: functional: pairing failure.Housing damaged broken inserts.Battery tray damage.Corrosion on the base plate.Minor scratches on the unit.Since the housing cannot be replaced, this unit is deemed unrepairable.No further testing or repairs on the unit will be conducted, and the unit will be placed into long term hold.The faulty parts was identified as the root cause for the device failure during the service evaluation.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot: manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.
 
Event Description
It was reported by the customer that preoperatively to an orthopedic surgery on (b)(6)2021, it was observed that the vapr vue wireless footswitch device did not pair with the machine.The batteries were changed and tried to re-pair but prompted error messages.During in-house engineering evaluation, it was determined that the housing was damaged with broken inserts on the device.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VAPR VUE WIRELESS FOOTSWITCH
Type of Device
FOOT-SWITCH, ELECTRICAL
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key13172333
MDR Text Key289427011
Report Number1221934-2022-00031
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K100638
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/05/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number227214
Device Lot Number792025
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/05/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown