Catalog Number SGC0702
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Low Blood Pressure/ Hypotension (1914); Tachycardia (2095); Pericardial Effusion (3271)
Event Date 12/13/2021
Event Type  Injury  
Manufacturer Narrative
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 4.The clip delivery system (cds) was advanced, and the clip was implanted successfully reducing mr to 1.When the steerable guide catheter was withdrawn to the right atrium, the patient's blood pressure dropped and tachycardia occurred.Echocardiogram showed a moderate to severe pericardial effusion.Medication was given and pericardiocentesis was performed successfully, and the patient was transferred to intensive critical care unit.The patient was reported to be doing well.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Based on the available information, a cause for the reported hypotension and pericardial effusion could not be determined.The reported tachycardia was due to reported hypotension.The reported patient effects of hypotension, pericardial effusion and tachycardia are listed in the instruction for use (ifu) as known possible complications associated with mitraclip procedures.The reported medical intervention, hospitalization and required medication was a result of case specific circumstances.There is no indication of a product issue with respect to manufacture, design or labeling.E1.Initial reporter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G4 STEERABLE GUIDING CATHETER
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13177105
MDR Text Key283319241
Report Number2024168-2022-00161
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeIS
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/06/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/15/2022
Device Catalogue NumberSGC0702
Device Lot Number10716R111
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/16/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
1 IMPLANTED MITRACLIP
Patient Outcome(s) Hospitalization; Required Intervention;
-
-