• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 SITE-RITE 8 ULTRASOUND; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 SITE-RITE 8 ULTRASOUND; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770501
Device Problems No Apparent Adverse Event (3189); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/17/2021
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
Per biomed, the unit locks up or freezes.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of system freezing was unconfirmed.The unit did not freeze or lock up during the testing of the unit.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.H3 other text : evaluation findings are in section h.11.
 
Event Description
Per biomed, the unit locks up or freezes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE-RITE 8 ULTRASOUND
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
DYMAX CORP. -2523003
110 marshall drive
warrendale PA 15086
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key13193617
MDR Text Key283493737
Report Number3006260740-2021-05582
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741098338
UDI-Public(01)00801741098338
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770501
Device Catalogue Number9770501
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/25/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-