Model Number 26633 |
Device Problems
Difficult to Advance (2920); Material Deformation (2976); Material Integrity Problem (2978)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/26/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that stent damage occurred.A 07f 10 090 placehit wallstent was advanced for treatment.During the procedure, the nose cone of the stent was observed to be out of shape under x-ray and it was difficult to advance the whole shaft.The physician removed the whole device and the procedure was completed using another of same device.There were no patient complications and the patient's status was stable.
|
|
Event Description
|
It was reported that stent damage occurred.A 07f 10 090 placehit wallstent was advanced for treatment.During the procedure, the nose cone of the stent was observed to be out of shape under x-ray and it was difficult to advance the whole shaft.The physician removed the whole device and the procedure was completed using another of same device.There were no patient complications and the patient's status was stable.It was further reported that there was no damaged on the stent struts.The 80-90% stenosed target lesion was located in the biliary duct.
|
|
Manufacturer Narrative
|
Device evaluated by mfr.: a placehit wallstent device was returned for analysis.The device was returned with the stent in the correct position on the delivery system.A visual and tactile examination identified damaged to the tip of the device.A visual and tactile examination identified a shaft kink 20mm distal from the distal end of the t connector.This concludes the product analysis.
|
|
Manufacturer Narrative
|
B5: updated event description.Device evaluated by mfr.: a placehit wallstent device was returned for analysis.The device was returned with the stent in the correct position on the delivery system.A visual and tactile examination identified damaged to the tip of the device.A visual and tactile examination identified a shaft kink 20mm distal from the distal end of the t connector.This concludes the product analysis.
|
|
Event Description
|
It was reported that stent damage occurred.A 07f 10 090 placehit wallstent was advanced for treatment.During the procedure, the nose cone of the stent was observed to be out of shape under x-ray and it was difficult to advance the whole shaft.The physician removed the whole device and the procedure was completed using another of same device.There were no patient complications and the patient's status was stable.It was further reported that there was no damaged on the stent struts.The 80-90% stenosed target lesion was located in the biliary duct.It was further reported that there was no contact with any other device until the stent was loaded onto the guidewire, then inserted into the sheath.After which the device was difficult to advance over the guide wire and it was removed.
|
|
Search Alerts/Recalls
|